Characterization and Commissioning of a Triple-GEM Detector
We present in detail the characterization and commissioning of a triple Gas Electron Multiplier (triple-GEM)detector. The response to various sources of radiation was studied, and its efficiency, gain, energy resolution,and temporal resolution were calculated. We found an energy resolution of 19.5%,...
Hlavní autoři: | , |
---|---|
Médium: | Online |
Jazyk: | spa |
Vydáno: |
Universidad Pedagógica y Tecnológica de Colombia
2023
|
Témata: | |
On-line přístup: | https://revistas.uptc.edu.co/index.php/ciencia_en_desarrollo/article/view/14191 |
Shrnutí: | We present in detail the characterization and commissioning of a triple Gas Electron Multiplier (triple-GEM)detector. The response to various sources of radiation was studied, and its efficiency, gain, energy resolution,and temporal resolution were calculated. We found an energy resolution of 19.5%, a temporal resolution of40 ns., and a maximum gain of 5×106. We conclude that the detector can be confidently used in furtherstudies of medical dosimetry, since it is working in the proportional mode. |
---|