Characterization and Commissioning of a Triple-GEM Detector
We present in detail the characterization and commissioning of a triple Gas Electron Multiplier (triple-GEM)detector. The response to various sources of radiation was studied, and its efficiency, gain, energy resolution,and temporal resolution were calculated. We found an energy resolution of 19.5%,...
Egile Nagusiak: | , |
---|---|
Formatua: | Online |
Hizkuntza: | spa |
Argitaratua: |
Universidad Pedagógica y Tecnológica de Colombia
2023
|
Gaiak: | |
Sarrera elektronikoa: | https://revistas.uptc.edu.co/index.php/ciencia_en_desarrollo/article/view/14191 |
Gaia: | We present in detail the characterization and commissioning of a triple Gas Electron Multiplier (triple-GEM)detector. The response to various sources of radiation was studied, and its efficiency, gain, energy resolution,and temporal resolution were calculated. We found an energy resolution of 19.5%, a temporal resolution of40 ns., and a maximum gain of 5×106. We conclude that the detector can be confidently used in furtherstudies of medical dosimetry, since it is working in the proportional mode. |
---|